Direct Determination of Mercury at 0.2 ppm Level by Flame AA Using a Thermospray Atomizer and a Circular Burner
作者:
J.W. Robinson,
D.S. Choi,
期刊:
Spectroscopy Letters
(Taylor Available online 1989)
卷期:
Volume 22,
issue 6
页码: 763-777
ISSN:0038-7010
年代: 1989
DOI:10.1080/00387018908053934
出版商: Taylor & Francis Group
数据来源: Taylor
摘要:
A circular flame atomizer system using a quartz T flame adapator has been developed for the direct determination of mercury. A thermospray nebulizer was used as a sample introduction device to the atomizer. This design gave a sensitivity (1% absorption) of 0.20 ppm and detection limits of 10 ppb at 1.0 mL/min solvent flow at 253.7 nm. The absolute sensitivity was 25 ng and absolute detection limits was 1.5 ng.