首页   按字顺浏览 期刊浏览 卷期浏览 Determination of Palladium by Graphite Furnace-Atomic Absorption Spectrometry with Vapr...
Determination of Palladium by Graphite Furnace-Atomic Absorption Spectrometry with Vaprization from a Platform

 

作者: János Fazakas,  

 

期刊: Analytical Letters  (Taylor Available online 1982)
卷期: Volume 15, issue 3  

页码: 245-265

 

ISSN:0003-2719

 

年代: 1982

 

DOI:10.1080/00032718208064380

 

出版商: Taylor & Francis Group

 

关键词: graphite furnace;palladium;atomic absorption;graphite platform;nonresonance lines.

 

数据来源: Taylor

 

摘要:

Vaporizing palladium from a graphite platform increases peak area and in some cases peak height sensitivity. The absorption pulse is considerably delayed so that vaporization takes place under isothermal conditions. Many vapor phase interferences are thus overcome. The relative sensitivity of the 340.5 nm non-resonance line exceeds that of the 276.3 nm resonance line. The mass of the platform as well as its contact surface with the graphite tube was found to exhibit a marked influence on peak height sensitivity. Calibration curves are more linear than with vaporization from the wall.

 

点击下载:  PDF (652KB)



返 回