首页   按字顺浏览 期刊浏览 卷期浏览 Spiral Chronograph for Measurement of Single Millisecond Time Intervals with Microsecon...
Spiral Chronograph for Measurement of Single Millisecond Time Intervals with Microsecond Accuracy

 

作者: R. J. Emrich,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1947)
卷期: Volume 18, issue 3  

页码: 150-157

 

ISSN:0034-6748

 

年代: 1947

 

DOI:10.1063/1.1740912

 

出版商: AIP

 

数据来源: AIP

 

摘要:

A chronograph is described which measures time intervals up to 2000 microseconds with an error not exceeding one microsecond. The chronograph traces a spiral on a cathode‐ray tube at a known constant angular velocity; the spiral trace starts with the reception of an electrical pulse marking the beginning of the time interval and stops with the reception of a second pulse marking the ending of the interval so that the number of revolutions in the spiral is a direct measure of the time interval. The cathode‐ray tube screen is of long persistence so that an immediate measurement of the time interval, as well as a photographic record, is obtained. Measurements may be made as rapidly as one every 30 seconds. A circuit used in conjunction with the chronograph selects the beginning of a slowly rising pulse, sharpens the pulse to one rising in about 0.2 microsecond, and blocks all subsequent pulses from the chronograph until reset manually.

 

点击下载:  PDF (555KB)