首页   按字顺浏览 期刊浏览 卷期浏览 A wide wavelength range spectrometer (1150–8000 A˚) for the RFX reversed fie...
A wide wavelength range spectrometer (1150–8000 A˚) for the RFX reversed field pinch experiment

 

作者: L. Carraro,   M. E. Puiatti,   P. Scarin,   M. Valisa,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1992)
卷期: Volume 63, issue 10  

页码: 5188-5190

 

ISSN:0034-6748

 

年代: 1992

 

DOI:10.1063/1.1143425

 

出版商: AIP

 

数据来源: AIP

 

摘要:

This paper describes the vacuum Czerny–Turner spectrometer with a 1.33 m focal length installed on the RFX fusion experiment. The optics are coated with MgF2and the image intensifier, protected by a vacuum sealed MgF2window, features a dual photocathode system, CsTe on one half and S20 on the other; thus the instrument covers efficiently the wide spectral range from the MgF2limit to the near infrared. Sensitivity at short wavelengths is limited by the reflectance of mirrors and the transmittance of the MgF2window. The detector is completed by a fast optical multichannel analyzer whose 1024 photodiodes can be read out in a time of 250 &mgr;s. The instrument can also be operated in a duochromator mode by deflecting the diffracted beam onto a second detector system, where two photomultipliers are applied to two exit slits, one fixed and the other remotely movable in such a way that the intensity ratios of two lines can be monitored for diagnostic purposes. Some examples of obtained spectra illustrating the instrument performance and comprising first results from the RFX plasma are presented.

 

点击下载:  PDF (409KB)



返 回