Compton scatter with polychromatic sources for lung densitometry

 

作者: James A. Hanson,   William E. Moore,   Melvin M. Figley,   Peter R. Duke,  

 

期刊: Medical Physics  (WILEY Available online 1998)
卷期: Volume 11, issue 5  

页码: 633-637

 

ISSN:0094-2405

 

年代: 1998

 

DOI:10.1118/1.595558

 

出版商: American Association of Physicists in Medicine

 

数据来源: WILEY

 

摘要:

A mobile lung densitometer using conventional x‐ray tubes, NaI detectors, and principles of two‐source, two‐detector Compton scattered densitometry, is described. The device is capable of one to two per second density measurements from a 45‐cm3volume with a precision of 5%. The expectedin vivoaccuracy (2%–3%) is determined by using an anthropomorphic phantom with replaceable lung inserts. The unintentional detection of multiple‐scattered x rays results in a small density‐dependent error. This error is predictable and relatively insensitive to differences in surrounding absorbers such as the chest wall. With this device, dynamicin vivodensitometry of the lung in the clinical laboratory and intensive care unit will be possible.

 