首页   按字顺浏览 期刊浏览 卷期浏览 Internal Solid Contact Electrode for the Determination of Clenbuterol in Pharmaceutical...
Internal Solid Contact Electrode for the Determination of Clenbuterol in Pharmaceutical Formulations and Human Urine

 

作者: Xian Xiang Sun,   Hassan Y. Aboul-Enein,  

 

期刊: Analytical Letters  (Taylor Available online 1999)
卷期: Volume 32, issue 6  

页码: 1143-1156

 

ISSN:0003-2719

 

年代: 1999

 

DOI:10.1080/00032719908542884

 

出版商: Taylor & Francis Group

 

关键词: Clenbuterol;Pharmaceuticals;Ion-selective electrode;Potentiometry

 

数据来源: Taylor

 

摘要:

A potentiometric clenbuterol-selective electrode was developed based on a conducting poly(aniline) (PA) film immobilized on a glassy carbon electrode surface and a poly(vinyl chloride)(PVC) sensing membrane containing an ion pair complex of clenbuterol(CB) with tetraphenylborate. The effects of various conditions for the electropolymerization of PA such as electrochemical method, conducting support, aniline concentration, polymer (PA) film thickness were investigated in relation to the slope and linearity of calibration graphs. The electrode exhibited a Nernstian response in the CB concentration range 1x10−1to 3.9x10−6M with a slope of 60.6mV/decade (25°C). The potentiometric selective coefficients against structurally similar drugs such as epinephrine and salbutamol were of the order of 1x10−3. The electrode was successfully applied to the determination of CB in drug formulations using the calibration curve technique and also in human urine.

 

点击下载:  PDF (492KB)



返 回