End-Column Amperometric Detection of 5-Fluorouracil by Capillary Zone Electrophoresis with a Carbon Fiber Microelectrode
作者:
Tianyan You,
Xiaoqing Wang,
Xiurong Yang,
Erkang Wang,
期刊:
Analytical Letters
(Taylor Available online 1999)
卷期:
Volume 32,
issue 6
页码: 1109-1119
ISSN:0003-2719
年代: 1999
DOI:10.1080/00032719908542881
出版商: Taylor & Francis Group
关键词: End-column;Amperometric detection;Capillary zone electrophoresis;5-Fluorouracil;Carbon fiber electrode.
数据来源: Taylor
摘要:
A rapid and sensitive detection method for the determination of 5-fluorouracil(5-FU) in real samples such as human urine and bovine serum albumin (BSA) was described. A carbon fiber microdisk electrode was used to perform end-column amperometric detection in capillary zone electrophoresis. The detection limit was as low as 2.5x10−7M and the wider linear range for the concentration was between 5x10−6and 1x10−4M with a correlation coefficient of 0.995.