Soft x‐ray imaging instrument for the Alcator A tokamak
作者:
R. Petrasso,
M. Gerassimenko,
F. H. Seguin,
J. Ting,
R. Krogstad,
P. Gauthier,
W. Hamilton,
A.T. Ramsey,
P. Burstein,
R. Granetz,
期刊:
Review of Scientific Instruments
(AIP Available online 1980)
卷期:
Volume 51,
issue 5
页码: 585-590
ISSN:0034-6748
年代: 1980
DOI:10.1063/1.1136258
出版商: AIP
数据来源: AIP
摘要:
We describe an instrument that images the Alcator A tokamak plasma in the soft x‐ray energy range with high temporal resolution. Radiation from a cross section of the plasma column is detected by an array of seventeen surface barrier detectors which provides 1 cm poloidal spatial resolution over a 16 cm field of view. The instrument’s nominal sensitivity lies in the spectral range from 0.27 to 25 keV; this range includes much of the radiation emitted by the Alcator A plasma, whose core temperature is typically in the 0.6 to 1.1 keV range. The frequency response of the detector‐electronics system extends from dc to 170 kHz. We discuss: (1) some design considerations and instrument calibrations, including the instrument’s measured response to 0.28, 1.5, 8.0, and 17.5 keV x‐rays and its theoretical response to a bremsstrahlung spectrum; (2) the detector‐electronics noise levels under various conditions; (3) the instrument’s response to an electrically modulated x‐ray source; (4) the detector x‐ray response as a function of bias voltage and temperature; (5) the detector linearity of response to 1.5 keV x‐rays; and (6) the stability of the detector x‐ray response despite numerous and prolonged vacuum breaks. Sample data, obtained from Alcator A discharges, are presented.