Argon triple point realization cryostat for platinum resistance long stem thermometers
作者:
J. Ancsin,
J. Phillips,
期刊:
Review of Scientific Instruments
(AIP Available online 1976)
卷期:
Volume 47,
issue 12
页码: 1519-1521
ISSN:0034-6748
年代: 1976
DOI:10.1063/1.1134567
出版商: AIP
数据来源: AIP
摘要:
A high‐precision realization cryostat was developed for long stem type platinum resistance thermometers at the argon triple point. Six liters of argon gas (at room temperature) are sufficient for calibration and if ultrahigh purity grade is used one may expect the results to yield better than ±1 mK accuracy.
点击下载:
PDF
(164KB)
返 回