首页   按字顺浏览 期刊浏览 卷期浏览 Argon triple point realization cryostat for platinum resistance long stem thermometers
Argon triple point realization cryostat for platinum resistance long stem thermometers

 

作者: J. Ancsin,   J. Phillips,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1976)
卷期: Volume 47, issue 12  

页码: 1519-1521

 

ISSN:0034-6748

 

年代: 1976

 

DOI:10.1063/1.1134567

 

出版商: AIP

 

数据来源: AIP

 

摘要:

A high‐precision realization cryostat was developed for long stem type platinum resistance thermometers at the argon triple point. Six liters of argon gas (at room temperature) are sufficient for calibration and if ultrahigh purity grade is used one may expect the results to yield better than ±1 mK accuracy.

 

点击下载:  PDF (164KB)



返 回