首页   按字顺浏览 期刊浏览 卷期浏览 A comparison of beam‐hardening artifacts in x‐ray computerized tomography with gadolini...
A comparison of beam‐hardening artifacts in x‐ray computerized tomography with gadolinium and iodine contrast agents

 

作者: Christopher Ruth,   Peter M. Joseph,  

 

期刊: Medical Physics  (WILEY Available online 1998)
卷期: Volume 22, issue 12  

页码: 1977-1982

 

ISSN:0094-2405

 

年代: 1998

 

DOI:10.1118/1.597495

 

出版商: American Association of Physicists in Medicine

 

数据来源: WILEY

 

摘要:

Lanthanide‐based compounds such as Gd‐DTPA are currently used as contrast agents in MRI. Recent experiments using CT and transmission radiography show that Gd agents can increase image contrast by up to a factor of 2 relative to more commonly used iodinated agents on an equi‐molar basis. It has also been suggested that beam hardening artifacts may be reduced with Gd. This hypothesis was experimentally tested on three different CT scanners using a circular water equivalent phantom with contrast filled tube inserts. It was found that the artifacts were a factor of 1.3–1.8 more pronounced with the iodinated contrast compared with Gd‐DTPA. A theoretical model which uses an experimentally derived photon energy spectrum is proposed which relates the strength of beam hardening artifacts to the variance (over the energy spectrum) of the attenuation coefficient of the contrast agent. This allows easy assessment of the relative magnitudes of the artifact for different contrast agents.

 

点击下载:  PDF (805KB)



返 回