Carbon Dioxide Supercritical Fluid Extraction with on-line Fluorescence Detection
作者:
D.C. Dunham,
M.S. Desmarais,
T. Breid,
J.W. Hills,
期刊:
Analytical Letters
(Taylor Available online 1995)
卷期:
Volume 28,
issue 14
页码: 2561-2574
ISSN:0003-2719
年代: 1995
DOI:10.1080/00032719508004035
出版商: Taylor & Francis Group
关键词: Supercritical fluid extraction;fluorescence spectroscopy
数据来源: Taylor
摘要:
An HPLC detector was modified to operate as a low pressure on-line supercritical fluid extraction detector. Spectra of CO2and fluorene are presented of pressures ranging from 85 to 150 atm. Both the intensity and emission wavelength of fluorene were effected by CO2pressure. As pressure increased, the emission intensity also increased and the emission shifted to a shorter wavelength. The intensity of CO2emission decreased by increasing pressure. The emission wavelength of CO2was unaffected by pressure. A detection limit for fluorene was found to be 5 μg per extraction. The minimal detectable mass was 25 μg (± 8% RSD) per extraction.
点击下载:
PDF (397KB)
返 回