首页   按字顺浏览 期刊浏览 卷期浏览 Extension of the application of the shorted quartz gauge
Extension of the application of the shorted quartz gauge

 

作者: Gideon Rosenberg,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1979)
卷期: Volume 50, issue 4  

页码: 442-444

 

ISSN:0034-6748

 

年代: 1979

 

DOI:10.1063/1.1135847

 

出版商: AIP

 

数据来源: AIP

 

摘要:

Shorted guard‐ring quartz gauges 12.7 mm in diameter and 3.2 mm thick were shock loaded by AL 6061 and copper impactors to stresses of 42 and 58 kilobars, respectively. Distortion of the current output was observed only at the 58 kilobar stress; at 42 kilobars the current profile conforms closely to the stress profile. Comparison of present data with previous studies demonstrates: (a) The present results lie on a smooth extension of an earlier calibration by Hayes and Gupta. (b) When only the current peak at the shock front is considered, the present data lie close to Graham’s curve for shunted gauges. An example of application of the shorted gauge at 40 kilobars to monitor a stress relaxing profile in LiF is also described.

 

点击下载:  PDF (211KB)



返 回