首页   按字顺浏览 期刊浏览 卷期浏览 Evaluation of dead‐time corrections in EDS systems
Evaluation of dead‐time corrections in EDS systems

 

作者: D. J. Bloomfield,   G. Love,   V. D. Scott,  

 

期刊: X‐Ray Spectrometry  (WILEY Available online 1983)
卷期: Volume 12, issue 1  

页码: 2-7

 

ISSN:0049-8246

 

年代: 1983

 

DOI:10.1002/xrs.1300120103

 

出版商: Wiley Subscription Services, Inc., A Wiley Company

 

数据来源: WILEY

 

摘要:

AbstractTwo common methods of applying dead‐time corrections in EDS systems, both of which are carried out internally in the electronics, are examined. The Barnhart method shows serious discrepancies when making dead‐time corrections and in certain cases no correction is made at all. The problem is associated with the reliance on the fast amplifier to count all x‐rays collected by the detector. In practice, however, the fast amplifier does not always succeed in carrying out this function since x‐rays with energies<1500 eV may fall below its threshold level. The problems become worse when a formvar window is substituted for beryllium since a greater proportion of low energy x‐rays can now reach the detector. The magnitude of the discrepancy is shown also to be a function of specimen and electron accelerating voltage. Similar experiments with an EDS system using the Covell method show that no such discrepanc

 

点击下载:  PDF (527KB)



返 回