首页   按字顺浏览 期刊浏览 卷期浏览 Reproducibility of a commercially available kit utilizing enzyme‐linked immunoso...
Reproducibility of a commercially available kit utilizing enzyme‐linked immunosorbent assay for determination of aflatoxin in peanut butter

 

作者: C. M. Ward,   M. R. A. Morgan,  

 

期刊: Food Additives & Contaminants  (Taylor Available online 1991)
卷期: Volume 8, issue 1  

页码: 9-15

 

ISSN:0265-203X

 

年代: 1991

 

DOI:10.1080/02652039109373951

 

出版商: Taylor & Francis Group

 

关键词: Aflatoxin;ELISA;reproducibility;peanut butter;kit

 

数据来源: Taylor

 

摘要:

The reproducibility of a commercially available kit utilising enzyme‐linked immunosorbent assay (ELISA) for determination of total aflatoxins has been analysed. Peanut butter samples contaminated with aflatoxin at three concentrations (mean 5.8, 11.7 and 24.8 ng/g) were analysed on different occasions using four kits. Coefficients of variation at all concentrations ranged from 2.6% to 13.6% within replicate, 0.3% to 16.2% within plate and 6.7% to 10.2% between kits. Standard curve precision profiles showed a minimum coefficient variation (CV) of 2–4% and a working range (below 10% CV) covering almost the entire standard curve.

 

点击下载:  PDF (374KB)



返 回