Monoclonal Antibody Biosensor for Antigen Monitoring
作者:
DianaL. Bush,
GarryA. Rechnitz,
期刊:
Analytical Letters
(Taylor Available online 1987)
卷期:
Volume 20,
issue 11
页码: 1781-1790
ISSN:0003-2719
年代: 1987
DOI:10.1080/00032718708082595
出版商: Taylor & Francis Group
关键词: Potentiometry;Membrane Electrodes;Dinitrophenol
数据来源: Taylor
摘要:
A new class of fully reversible biosensors, utilizing a layer of monoclonal antibodies trapped at the sensing surface of an immobilized-antigen membrane, is shown to be suited for continuous monitoring of antigen levels. The model sensor described gives a potentiometric response to the hapten 2,4-dinitrophenol (DNP) at micromolar concentration levels through competitive interaction with monoclonal anti-DNP antibodies held within the sensor tip. The proposed design yields a reusable, reagentless antigen sensor with good analytical precision and an operating lifetime of at least 17 days.