首页   按字顺浏览 期刊浏览 卷期浏览 Gamma‐Ray Attenuation Factors for Angular Correlation and Angular Distribution M...
Gamma‐Ray Attenuation Factors for Angular Correlation and Angular Distribution Measurements

 

作者: D. W. Glasgow,   L. W. Coleman,   L. Schecter,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1961)
卷期: Volume 32, issue 6  

页码: 683-684

 

ISSN:0034-6748

 

年代: 1961

 

DOI:10.1063/1.1717468

 

出版商: AIP

 

数据来源: AIP

 

摘要:

The angular resolution and azimuthal symmetry of an unshielded NaI scintillation counter was measured for gamma rays between 0.5 and 2.8 Mev in order to estimate correction factors needed in determining angular correlation functions. The results are in agreement with the corrections calculated by Rose, and extended by Stanford and Rivers, for similar scintillation crystal geometry.

 

点击下载:  PDF (148KB)



返 回