Flow Injection Analysis of Glucose by Fiber Optic Chemiluminescence Measurement
作者:
AhmadA. Suleiman,
RhodoraL. Villarta,
GeorgeG. Guilbault,
期刊:
Analytical Letters
(Taylor Available online 1993)
卷期:
Volume 26,
issue 7
页码: 1493-1503
ISSN:0003-2719
年代: 1993
DOI:10.1080/00032719308017427
出版商: Taylor & Francis Group
关键词: glucose;optrode;chemiluminescence;flow-injection analysis
数据来源: Taylor
摘要:
A flow injection system was developed for the determination of glucose based on fiber optic chemiluminescence measurements. The hydrogen peroxide produced from the glucose oxidase catalyzed reaction was quantified by measuring the intensity of luminol chemiluminescence in the presence of excess ferricyanide. Glucose oxidase was immobilized on aminopropyl glass using glutaraldehyde and packed in a reactor. The chemiluminescence was transported to the detector using an optical fiber. The system responds linearly to glucose in the concentration range 0.20 mM up to 2.5 mM, with a relative precision of 5%. Several fruit juices were analyzed for their glucose content and the results were compared with a standard AOAC procedure.