Determination of Enzymatic Activities Based on an Optical Flow-Through p-Nitrophenol Sensor
作者:
J.M. Fernández-Romero,
M.D. Luque de Castro,
M. Valcárcel,
期刊:
Analytical Letters
(Taylor Available online 1993)
卷期:
Volume 26,
issue 9
页码: 1847-1866
ISSN:0003-2719
年代: 1993
DOI:10.1080/00032719308017435
出版商: Taylor & Francis Group
关键词: flow-through (bio)sensor;p-nitrophenol;β-D-glucuronidase;serum
数据来源: Taylor
摘要:
An optical flow-through biosensor based on the transient retention of p-nitrophenol on an ion-exchange support packed in the flow cell is presented. The approach was applied to the determination of the activity of some enzymes which catalyze the conversion of p-nitrophenyl-derivatives into p-nitrophenol. The method was first developed for determination of p-nitrophenol yielding a linear range between 0.1–5 μg/mL (r2= 0.9922, r.s.d.% less than 2.5), then, it was also applied to the determination of β-D-glucuronidase activity in serum, with a linear range between 0.1–20 U/L (r2= 0.9976, r.s.d.% less than 3.0), and a sampling frequency of 20 h−1. The application of the method to the determination of the enzyme activity in serum samples provided results consistent with those obtained by a conventional method and recoveries within 95–104%.
点击下载:
PDF (468KB)