首页   按字顺浏览 期刊浏览 卷期浏览 Measurement of dissolved carbon dioxide
Measurement of dissolved carbon dioxide

 

作者: J. S. Alford Jr.,  

 

期刊: Canadian Journal of Microbiology  (NRC Available online 1976)
卷期: Volume 22, issue 1  

页码: 52-56

 

ISSN:0008-4166

 

年代: 1976

 

DOI:10.1139/m76-007

 

出版商: NRC Research Press

 

数据来源: NRC

 

摘要:

Several probes for measuring dissolved carbon dioxide (CO2) concentration were installed in a 68-litre fermentor and their effectiveness compared. Submerged silastic rubber tubing gave reproducible results over a wide range of operating conditions and was generally superior to all other probes evaluated.The silastic rubber probe was used to compare the partial pressure of CO2in viscous fermentation media with that in the fermentor exhaust gas. No significant difference was found.Results show that determination of the CO2partial pressure in the exhaust gas gives an excellent approximation of the partial pressure of dissolved CO2in the liquid medium, eliminating the need for measurement of CO2concentration in the broth.

 