首页   按字顺浏览 期刊浏览 卷期浏览 Increased ion intensity and reliability of the Stockholm electron beam ion sourcea)
Increased ion intensity and reliability of the Stockholm electron beam ion sourcea)

 

作者: E. Beebe,   L. Liljeby,   A. Pikin,   M. Bjo¨rkhage,   A˚. Engstro¨m,   A. Paal,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1994)
卷期: Volume 65, issue 5  

页码: 1718-1722

 

ISSN:0034-6748

 

年代: 1994

 

DOI:10.1063/1.1144865

 

出版商: AIP

 

数据来源: AIP

 

摘要:

The electron beam ion source, CRYSIS, produces highly charged ions for injection into the heavy ion storage ring—CRYRING at MSL, as well as low energy atomic physics experiments and the Stockholm–Mainz Penning trap recently installed at MSL. CRYSIS has produced ions up to Ar18+and136Xe52+. Pulsed beams of Ar13+ions 60 &mgr;s in duration have been injected into CRYRING via an RFQ and ions of charge up to136Xe44+have been used in atomic physics experiments with pulse duration 10–250 ms. A vacuum separation of the cryostat and ionization volumes has been made. Temperature control and measurement of internal electrodes have increased the gas injection efficiency and reduced the memory effect associated with a cryogenic EBIS. External ion injection has been added as an alternative to neutral gas injection for introducing the species to be ionized to high charge states. Monitoring of the radio frequency noise signal with a spectrum analyzer has aided in the propagation of quiet, high current (450 mA) dc electron beams. These quiet electron beams have been used to produce extracted ion pulses of higher intensities than in previous operation.

 

点击下载:  PDF (651KB)