Routine plasma colloid osmotic pressure measurements
作者:
MAX WEIL,
MARTIN MORISSETTE,
SYBIL MICHAELS,
JOE BISERA,
EDWARD BOYCKS,
HERBERT SHUBIN,
EDWIN JACOBSON,
期刊:
Critical Care Medicine
(OVID Available online 1974)
卷期:
Volume 2,
issue 5
页码: 229-234
ISSN:0090-3493
年代: 1974
出版商: OVID
数据来源: OVID
摘要:
A method for routine determination of colloid osmotic pressure (COP) in the clinical laboratory is described. A transducer membrane system is utilized in which the colloid osmotic pressure (COP) of fresh or refrigerated plasma or albumin is compared to that of isotonic saline. The measurement is completed within an average period of less than four minutes. Duplicate measurements demonstrated high reproducibility (r = >.95) and reference measurements obtained over a period of ten months were consistently within a range of 1 torr. Heparin in excess of 200 units/ml delayed equilibration. Values obtained in normal ambulatory and supine subjects conform closely to those previously reported.