首页   按字顺浏览 期刊浏览 卷期浏览 A Mass Spectrometer for the Simultaneous Measurement of the Neutral and the Ion Composi...
A Mass Spectrometer for the Simultaneous Measurement of the Neutral and the Ion Composition of the Upper Atmosphere

 

作者: H. Balsiger,   P. Eberhardt,   J. Geiss,   E. Kopp,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1971)
卷期: Volume 42, issue 4  

页码: 475-476

 

ISSN:0034-6748

 

年代: 1971

 

DOI:10.1063/1.1685134

 

出版商: AIP

 

数据来源: AIP

 

摘要:

A mass spectrometer is described which measures the neutral and the ionized components of the upper atmosphere simultaneously (delay ∼20 msec). Its 1% mass resolution is 60 and its sensitivity is sufficient to measure small ion concentrations in the presence of high neutral gas densities. During rocket flights the shift in the mass scale between the ion peak and the corresponding neutral peak is a measure for the rocket skin potential.

 