首页   按字顺浏览 期刊浏览 卷期浏览 Detection of aortic insufficiency by pulse doppler echocardiography
Detection of aortic insufficiency by pulse doppler echocardiography

 

作者: Jonathan M. Ward,   Donald W. Baker,   Simeon A. Rubenstein,   Steve L. Johnson,  

 

期刊: Journal of Clinical Ultrasound  (WILEY Available online 1977)
卷期: Volume 5, issue 1  

页码: 5-10

 

ISSN:0091-2751

 

年代: 1977

 

DOI:10.1002/jcu.1870050103

 

出版商: Wiley Subscription Services, Inc., A Wiley Company

 

数据来源: WILEY

 

摘要:

AbstractSixty‐five patients were studied for the presence of aortic insufficiency by auscultation, conventional M mode echocardiography (echo), range‐gated pulse Doppler flowmeter (Doppler), and supravalvular aortic angiography. Aortic insufficiency was demonstrated angiographically in 49 patients, of which eight (16 per cent) were missed by auscultation. Echo missed 20 (41 per cent), and Doppler missed seven (14 per cent). Only two (4 per cent) were missed by auscultation and Doppler combined. All methods were highly specific, but the combination of auscultation and Doppler was the most sensitive noninvasive method for detecting aortic insufficie

 