Millimicrosecond Timing with Large Scintillation Detectors
作者:
J. W. Keuffel,
W. H. Sandmann,
R. K. Stitt,
期刊:
Review of Scientific Instruments
(AIP Available online 1955)
卷期:
Volume 26,
issue 12
页码: 1191-1195
ISSN:0034-6748
年代: 1955
DOI:10.1063/1.1715219
出版商: AIP
数据来源: AIP
摘要:
The timing precision attainable with large liquid scintillation detectors has been studied. Such detectors have slow rise‐times (∼10−8sec) compared with those used in conventional nuclear physics timing experiments. The rise‐time is a result of photon time of flight and transit time in the larger, slower photomultipliers (such as the Dumont 6364) which must be used for good light collection. However, timing errors due to rise‐time may be largely overcome when dealing with large, statistically smooth pulses such as are encountered in cosmic‐ray experiments with unstable particles. A simple correction based on the pulse height is applied to each time measurement. Using this technique a timing error curve is obtained with a standard deviation of 1.6 m&mgr;sec for a large detector using a Dumont 6364.