首页   按字顺浏览 期刊浏览 卷期浏览 High‐sensitivity helium resonance magnetometers
High‐sensitivity helium resonance magnetometers

 

作者: Douglas D. McGregor,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1987)
卷期: Volume 58, issue 6  

页码: 1067-1076

 

ISSN:0034-6748

 

年代: 1987

 

DOI:10.1063/1.1139609

 

出版商: AIP

 

数据来源: AIP

 

摘要:

A number of techniques to improve the sensitivity of helium resonance magnetometers is explored. Using a mathematical model of the helium cell discharge and optical pumping process, calculations indicate that substantial gains in sensitivity of a He4magnetometer are possible. The improvements are obtained by use of larger cells and the replacement of the lamp with a laser as the source of pump radiation. A large cell (2 in. diameter×6 in. long) magnetometer with lamp pumping is calculated to have a sensitivity of 3×10−4&ggr;/(Hz)1/2, a value which has been experimentally verified. The projected sensitivity of a laser‐pumped magnetometer varies between 2×10−5and 4×10−6&ggr;/(Hz)1/2, depending on cell size.

 

点击下载:  PDF (727KB)