首页   按字顺浏览 期刊浏览 卷期浏览 False coronary dissection with the new Monorail angioplasty balloon catheter
False coronary dissection with the new Monorail angioplasty balloon catheter

 

作者: Enrique Esplugas,   Angel R. Cequier,   Xavier Sabaté,   Francisco Jara,  

 

期刊: Catheterization and Cardiovascular Diagnosis  (WILEY Available online 1990)
卷期: Volume 19, issue 1  

页码: 30-33

 

ISSN:0098-6569

 

年代: 1990

 

DOI:10.1002/ccd.1810190109

 

出版商: Wiley Subscription Services, Inc., A Wiley Company

 

关键词: coronary angioplasty;cardiac catheterization;complications

 

数据来源: WILEY

 

摘要:

AbstractDuring percutaneous transluminal coronary angioplasty, the appearance of persistent staining in the vessel by contrast media suggests coronary dissection. We report seven patients in whom a false image of severe coronary dissection was observed during angioplasty performed with the new Monorail balloon catheter. This image emerges at the moment of balloon inflation, is distally located to the balloon, and disappears with balloon catheter deflation. No complications were associated with the appearance of this image.

 

点击下载:  PDF (371KB)



返 回