首页   按字顺浏览 期刊浏览 卷期浏览 Modification of an Intrauterine Shunt Catheter to Reduce the Incidence of Spontaneous D...
Modification of an Intrauterine Shunt Catheter to Reduce the Incidence of Spontaneous Displacement

 

作者: DOUGLAS RICHARDS,   GREGORY LOCKSMITH,   MICHAEL YANCEY,  

 

期刊: Obstetrics & Gynecology  (OVID Available online 1994)
卷期: Volume 84, issue 1  

页码: 52-54

 

ISSN:0029-7844

 

年代: 1994

 

出版商: OVID

 

数据来源: OVID

 

摘要:

ObjectiveTo determine which of four types of catheters would be the least likely to become displaced spontaneously when employed as a fetoamniotic shunt, by testing the force required to remove the catheters from a lamb's bladder.MethodsThe catheter designs tested were unfeathered straight, unfeathered pig-tailed, superficially feathered, and deeply feathered. The catheters were placed percutaneously into the urinary bladders of recently euthanized newborn lambs. The force required to withdraw the catheters was determined, employing nine measurements for each catheter design.ResultsThe mean force required to remove the straight catheter was 12 g (95% confidence interval [CI] 2–21 g), the pig-tailed catheter 22 g (95% CI 13–32 g), and the superficially feathered catheter 149 g (95% CI 139–160 g). Extracting the deeply feathered catheter consistently required a force greater than 300 g, which was the upper limit of detection for our instruments.conclusionFeathering the catheter increases the force required to extract it from a fetal viscus. We believe that use of feathered catheters will significantly reduce the incidence of unwanted spontaneous displacement.

 

点击下载:  PDF (200KB)