首页   按字顺浏览 期刊浏览 卷期浏览 Quantification of Muscle Tissue Magnesium and Potassium Using Atomic Absorption Spectro...
Quantification of Muscle Tissue Magnesium and Potassium Using Atomic Absorption Spectrometry

 

作者: K.I. Al-Khamis,   B.M. Al-Hadiyah,   S.A. Bawazir,   O.M. Ibrahim,   M.J. Al-Yamani,  

 

期刊: Analytical Letters  (Taylor Available online 1995)
卷期: Volume 28, issue 6  

页码: 1033-1053

 

ISSN:0003-2719

 

年代: 1995

 

DOI:10.1080/00032719508002677

 

出版商: Taylor & Francis Group

 

关键词: Determination of magnesium;potassium;cardiac muscle;skeletal muscle;muscle tissue;atomic absorption spectrometry;AAS;Mg2+;K+

 

数据来源: Taylor

 

摘要:

Flame atomic absorption spectrometry (AAS) has been used to measure magnesium (Mg2+) and potassium (K+) contents of wistar rat myocardium and skeletal muscle. Ten to twenty mg of tissue was homogenized in 10 mM hydrochloric acid, centrifuged, and the clear supernatant was diluted three-fold in deionized distilled water before determination of Mg2+and K+concentrations. The calibration curves of both ions were linear from 0.25 to 2.0 mg/L for Mg2+and from 1.5 to 15.0 mg/L for K+. The mean values (±SD) of Mg2+and K+contents for the analyzed cardiac muscle were 371.97±12.16 mg/Kg of wet tissue with a coefficient of variation (CV) of 3.3%, and 4571.02±359.74 mg/Kg with CV of 7.9%, respectively. Those for various skeletal muscles varied between 323.35-420.60 mg/Kg Mg2+, with CV ranging from 3.2% to 8.7%, and 4883.84-8156.68 mg/Kg K+, with CV ranging from 4.4% to 10.1%. The within-day CV were 2.8% to 3.8% for Mg2+and 3.2% to 5.9% for K+, whereas day-to-day CV were 6.2% to 10.2% for Mg2+and 8.7% to 10.5% for K+. Recoveries were 92.7% to 104.9% for Mg2+, and 93.2% to 106.1% for K+. Values for Mg2+ obtained by the proposed method were significantly different from those obtained by a published tissue digestion method, those for K+were not significantly different.

 

点击下载:  PDF (574KB)



返 回