首页   按字顺浏览 期刊浏览 卷期浏览 Determination of Trace Manganese by Adsorption Voltammetry
Determination of Trace Manganese by Adsorption Voltammetry

 

作者: Lizeng Wang,   Chengsong Ma,   Xiaoli Zhang,   Junguo Wang,  

 

期刊: Analytical Letters  (Taylor Available online 1993)
卷期: Volume 26, issue 8  

页码: 1711-1717

 

ISSN:0003-2719

 

年代: 1993

 

DOI:10.1080/00032719308021491

 

出版商: Taylor & Francis Group

 

关键词: Manganese;Adsorptive stripping voltammetry;Strawberry

 

数据来源: Taylor

 

摘要:

In a solution containing 0.005M borax medium (PH= 8− 10). 0.03% ascorbic acid and 0.001% gelatin, a fine sensitive adsorptive reduction peak of manganese appeared at −1.53v (vs.SCE) on a hanging mercury electrode by fast speed scanning voltammetry. The derivative peak current is directly proportional to the concentration of manganese in the range from 1.0×10−8M to 2.0×10−6M. The detection limit is 3.0×10−9M. Using this method, we have successfuly determined tracesof Mn in water and strawberry samples.

 