首页   按字顺浏览 期刊浏览 卷期浏览 Adiabatic Solution Calorimeter
Adiabatic Solution Calorimeter

 

作者: G. C. Benson,   E. D. Goddard,   C. A. J. Hoeve,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1956)
卷期: Volume 27, issue 9  

页码: 725-727

 

ISSN:0034-6748

 

年代: 1956

 

DOI:10.1063/1.1715682

 

出版商: AIP

 

数据来源: AIP

 

摘要:

The solution calorimeter described previously [Rev. Sci. Instr.26, 477 (1955)] has been modified by using an electrically heated shield for the adiabatic control. In this way the response time has been decreased by a factor of ten which greatly facilitates operation of the control. Under steady conditions the fluctuation in temperature between the calorimeter and its surroundings has been reduced from ±0.002 to ±0.001°C max. The calorimeter has a working range of 10° to 50°C.

 