首页   按字顺浏览 期刊浏览 卷期浏览 A Wide—Band Audio Phasemeter
A Wide—Band Audio Phasemeter

 

作者: John R. Ragazzini,   Lotfi A. Zadeh,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1950)
卷期: Volume 21, issue 2  

页码: 145-148

 

ISSN:0034-6748

 

年代: 1950

 

DOI:10.1063/1.1745513

 

出版商: AIP

 

数据来源: AIP

 

摘要:

The problem of measuring the phase angle between two sinusoidal voltages is an important one in laboratory procedure. Phasemeters which can make this measurement accurately over a substantial frequency range have heretofore been bulky and relatively stationary. This paper describes a circuit which results in a phasemeter which is accurate to about 30 minutes, has a dial adjustment which is independent of frequency, operates over a range of from 20 c.p.s. to 100 kc and is relatively small and inexpensive. The principle used is to match the phase shift between the unknown voltages with a calibrated phase shift in the meter channel. This phase equalization is observed on a standard oscillograph which is used only as a comparison instrument. The oscillograph amplifier phase shifts are balanced out with an uncalibrated phase shifter cascaded in the meter channel and do not affect the measurement.

 

点击下载:  PDF (299KB)



返 回