首页   按字顺浏览 期刊浏览 卷期浏览 A PC‐based instrument of a modified differential normal pulse voltammetry with b...
A PC‐based instrument of a modified differential normal pulse voltammetry with background current correction technique

 

作者: B. Y. Liao,   M. S. Young,   C. Y. Wang,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1994)
卷期: Volume 65, issue 5  

页码: 1679-1685

 

ISSN:0034-6748

 

年代: 1994

 

DOI:10.1063/1.1144859

 

出版商: AIP

 

数据来源: AIP

 

摘要:

Instead of the fixed pulse durations in conventional differential normal pulse voltammetry, a PC‐based instrument using adaptable pulse duration vectors for background current correction has been proposed and implemented. With this PC‐based instrument, a sequence of pulse duration vectors is determined in an initial blank solution voltammetry by assigning a decay percentage constant to the charging current. These vectors are then used in the subsequent test solution in order to obtain a pure oxidation current response in a resulting voltammogram with its corresponding features. Experimental results are presented to show the effectiveness and advantages of the PC‐based instrument for the measurements of dopamine in a phosphate buffered solution.

 

点击下载:  PDF (760KB)



返 回