首页   按字顺浏览 期刊浏览 卷期浏览 Observations on the use of a thermal conductivity cell to measure the para hydrogen con...
Observations on the use of a thermal conductivity cell to measure the para hydrogen concentration in a mixture of para and ortho hydrogen gas

 

作者: T. W. Bradshaw,   J. O. W. Norris,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1987)
卷期: Volume 58, issue 1  

页码: 83-85

 

ISSN:0034-6748

 

年代: 1987

 

DOI:10.1063/1.1139519

 

出版商: AIP

 

数据来源: AIP

 

摘要:

This paper describes the use of a thermal conductivity cell to measure the para hydrogen concentration in a mixture of para and ortho hydrogen gas. The device was calibrated over its entire range and the output compared with the theoretically expected form. The output of this device had been previously assumed to be linear, this assumption can lead to systematic errors of up to 5.6% in the determination of the para hydrogen concentration. A simple calibration procedure is described.

 