首页   按字顺浏览 期刊浏览 卷期浏览 Vibratory Sample Magnetometry of Middle Ear Prostheses and Manufacturing Materials
Vibratory Sample Magnetometry of Middle Ear Prostheses and Manufacturing Materials

 

作者: Mark Syms,   Derrick Peterman,  

 

期刊: Otology & Neurotology  (OVID Available online 2001)
卷期: Volume 22, issue 4  

页码: 487-491

 

ISSN:1531-7129

 

年代: 2001

 

出版商: OVID

 

关键词: Middle ear prostheses;Magnetic resonance imaging

 

数据来源: OVID

 

摘要:

ObjectiveTo assess the magnetic properties of stapes prostheses and manufacturing materials using a vibratory sample magnetometer (VSM).Data SourcesVSM was performed on 16 samples, including ferromagnetic 420F stainless steel, with an LDJ Model 9600 VSM in accordance with American Society for Testing and Materials standard A894.ResultsA VSM measures the magnetic dipole moment of a sample in a magnetic field. The magnetic field is swept over a range of magnetic fields, and the magnetic dipole moment is plotted as a function of field. The prostheses made of 316L stainless steel previously found to be ferromagnetic had the highest specific magnetic moments. The specific magnetic moments ranged from 0.023 electromagnetic units (emu)/g to 156 emu/g. The samples made with 316L stainless steel, which is used in otologic implants, were significantly less magnetic than was the 420F stainless steel.ConclusionVSM demonstrates that prostheses made with 316L stainless steel are relatively nonferromagnetic compared with 420F stainless steel. However, none of the implants are nonmagnetic. The torque and linear force on the prosthesis is a concern. The safety of performing MRI on patients with these implants needs to be further assessed.

 

点击下载:  PDF (53KB)