首页   按字顺浏览 期刊浏览 卷期浏览 Temperature Effects in Nuclear Magnetic Resonance Spectroscopy
Temperature Effects in Nuclear Magnetic Resonance Spectroscopy

 

作者: George Slomp,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1959)
卷期: Volume 30, issue 11  

页码: 1024-1027

 

ISSN:0034-6748

 

年代: 1959

 

DOI:10.1063/1.1716411

 

出版商: AIP

 

数据来源: AIP

 

摘要:

The effect of temperature on the stability of the Varian V‐4300–2 NMR spectrometer has been investigated. The spectrometer was most sensitive to changes in room‐air temperature. A change of 0.3°C per minute caused a spectrometer drift of about 20 cps per minute. A change of 0.1°C per minute in temperature of magnet cooling water caused a spectrometer drift of 1 cps per minute. Enclosing the magnet served to minimize effects of air temperature changes but amplified the effects of water temperature changes. Satisfactory control was attained by running the refrigeration units continuously and regulating the temperatures by controlling the suction pressure on the evaporators. Under these conditions spectra were reproducible to within ½ cps.

 