Concurrent Failure of Active and Redundant Ventricular Epicardial Electrodes in Children

 

作者: GERALD A. SERWER,   MACDONALD DICK,   KAREN UZARK,   WILLIAM A. SCOTT,   EDWARD L. BOVE,  

 

期刊: Journal of Interventional Cardiology  (WILEY Available online 1989)
卷期: Volume 2, issue 1  

页码: 43-47

 

ISSN:0896-4327

 

年代: 1989

 

DOI:10.1111/j.1540-8183.1989.tb00752.x

 

出版商: Blackwell Publishing Ltd

 

数据来源: WILEY

 

摘要:

Implanting an unused or redundant epicardial ventricular pacing electrode in children at the time of initial electrode placement for later use is a common practice. To investigate the ultimate utilization of this redundant electrode, the records of 38 children who had undergone implantation of 47 redundant epicardial electrodes between 1974 and 1987 were reviewed. As the primary electrode remains functional, 12 electrodes have not been in‐vasively tested or used; 11 were in nine children lost to follow‐up. The remaining 24 redundant electrodes implanted in 16 children, (ages 1 day to 18 years, median 5 years, at implantation) together with the utilized electrode were evaluated at the time of pacing system revision. At implant, all electrodes had thresholds<1.4 volts at a pulse duration of0.5 msec. At follow‐up 0.2 to 11.7 years (mean 3.7 years) after implant, an electrode was considered functional if the threshold at 0.5 msec pulse duration was<2.5 volts, and 14 of 24 redundant electrodes were nonfunctional. Of the ten functional redundant electrodes, the active electrode was functional in seven cases. In only three of 24 cases was the redundant electrode functional when the active electrode was nonfunctional. In all cases, when the active electrode exhibited exit block, the redundant electrode also exhibited exit block. Two of five redundant electrodes paired with a fractured active electrode were fractured. Two other redundant electrodes were fractured when the active electrode was functional. The state of the redundant and the active electrode was the same in 80% of the cases, P<0.03. Of the 36 redundant electrodes implanted for which follow‐up data is available, 33 had not been used by the end of the study period. (J Interven Cardiol 1

 

点击下载:  PDF (417KB)



返 回