A liquid‐helium‐cooled absolute reference cold load for long‐wavelength radiometric calibration

 

作者: Marc Bensadoun,   Chris Witebsky,   George Smoot,   Giovanni De Amici,   Al Kogut,   Steve Levin,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1992)
卷期: Volume 63, issue 10  

页码: 4377-4389

 

ISSN:0034-6748

 

年代: 1992

 

DOI:10.1063/1.1143738

 

出版商: AIP

 

数据来源: AIP

 

摘要:

We describe a large (78 cm) diameter liquid‐helium‐cooled blackbody absolute reference cold load for the calibration of microwave radiometers. The load provides an absolute calibration near the liquid‐helium (LHe) boiling point, with total uncertainty in the radiometric temperature of less than 30 mK over the 2.5–23‐cm wavelength (12–1.3 GHz) operating range. Emission from those parts of the cold load not immersed in LHe is ≤25 mK and the reflection coefficient is ≤3.5×10−4. This cold load has been used at several wavelengths at the South Pole, Antarctica and at the White Mountain Research Station, California to calibrate spectral measurements of the cosmic microwave background radiation. For the instruments operated at 20‐, 12‐, 7.9‐, and 4.0‐cm wavelength at the South Pole, the total corrections to the LHe boiling‐point temperature (∼3.8 K) were 48±23, 18±10, 10±18, and 15±16 mK, respectively. In operation, the average LHe loss rate was ≤4.4l/h, allowing day‐long periods of operation without a LHe fill. The boiloff rate is not strongly dependent on the radiative load at the aperture, yielding very stable operation and radiometric performance. Design considerations, radiometric and thermal performance, and operational aspects are discussed. A comparison with other LHe‐cooled reference loads including the predecessor of this cold load is given.

 

点击下载:  PDF (1739KB)



返 回