首页   按字顺浏览 期刊浏览 卷期浏览 Chronic Monitoring of Intracranial Pressure Using an in Vivo Calibrating SensorExperien...
Chronic Monitoring of Intracranial Pressure Using an in Vivo Calibrating SensorExperience in Patients with Pseudotumor Cerebri

 

作者: Paul Cooper,   Sarah Moody,   Frederick Sklar,  

 

期刊: Neurosurgery  (OVID Available online 1979)
卷期: Volume 5, issue 6  

页码: 666-670

 

ISSN:0148-396X

 

年代: 1979

 

出版商: OVID

 

关键词: Intracranial pressure;Long term monitoring;Pseudotumor cerebri

 

数据来源: OVID

 

摘要:

&NA;A number of reliable methods are available for the short term monitoring of intracranial pressure (ICP). However, the danger of infection and the need for external connections make the use of these methods for extended periods inappropriate. Most previously described devices for chronic monitoring of ICP are of limited value because of uncertain drift from zeropoint readings. The Hittman‐Meyer ICP sensor approaches the ideal for a long term monitoring device. It is fully implantable and is accurate over a wide range of pressures and for extended periods. It is designed so that in vivo calibration may be performed in a noninvasive fashion. We report the use of this device in eight patients with pseudotumor cerebri. It has functioned accurately for periods ranging up to 14 months after implantation and has proven to be a valuable aid in the management of these patients. (Neurosurgery,5: 666‐670, 1979)

 

点击下载:  PDF (2549KB)



返 回