首页   按字顺浏览 期刊浏览 卷期浏览 Comparison of some glass thermometers at low temperatures in a high magnetic field
Comparison of some glass thermometers at low temperatures in a high magnetic field

 

作者: S. A. J. Wiegers,   R. Jochemsen,   C. C. Kranenburg,   G. Frossati,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1987)
卷期: Volume 58, issue 12  

页码: 2274-2278

 

ISSN:0034-6748

 

年代: 1987

 

DOI:10.1063/1.1139335

 

出版商: AIP

 

数据来源: AIP

 

摘要:

We have compared the low‐frequency capacitance of glass thermometers of various shapes, made of very pure SiO2doped with ≊1200 ppm OH−, in the temperature range of 4–100 mK in a magnetic field of 9 T. The construction of the electrodes, the way the contacts are made, and the connection of the leads or cables influence both the voltage and the field dependence of the capacitance and of the loss. We have been able to construct capacitive glass thermometers of which the capacitance is not shifted in a field of 9 T at temperatures down to 4 mK, within our temperature accuracy of ≊5%. This has been determined by comparison with a cerium magnesium nitrate (CMN) thermometer in a field‐free region, and by comparing ultrasound measurements versus the thermometer capacitance with and without a magnetic field.

 

点击下载:  PDF (512KB)



返 回