首页   按字顺浏览 期刊浏览 卷期浏览 The Application of an Enzyme Electrode for Blood Glucose Determination in the Automated...
The Application of an Enzyme Electrode for Blood Glucose Determination in the Automated Flow System

 

作者: K. Bertermann,   P. Elze,   F. Scheller,   D. Pfeiffer,   M. Janchen,  

 

期刊: Analytical Letters  (Taylor Available online 1982)
卷期: Volume 15, issue 4  

页码: 397-404

 

ISSN:0003-2719

 

年代: 1982

 

DOI:10.1080/00032718208064391

 

出版商: Taylor & Francis Group

 

关键词: Determination of Blood Glucose;Enzyme Electrode;Automated Flow System

 

数据来源: Taylor

 

摘要:

An enzyme electrode containing immobilized GOD was utilized in a continuous flow system for the automatical determination of glucose. The advantages of kinetic measurement of H2O2are used. Characteristic parameters of device and results in routine are shown. Particular advantages are: low requirement on material (20 μl blood, < l μg GOD/measurement) and time (60 samples/h at 15 sec response time) as well as sufficient precision and accuracy (S % - serial: < 2 %).

 