Quartz Helix Magnetic Susceptibility Balance Using the Curie‐Cheneveau Principle
作者:
F. E. Senftle,
M. D. Lee,
A. A. Monkewicz,
J. W. Mayo,
Titus Pankey,
期刊:
Review of Scientific Instruments
(AIP Available online 1958)
卷期:
Volume 29,
issue 5
页码: 429-432
ISSN:0034-6748
年代: 1958
DOI:10.1063/1.1716216
出版商: AIP
数据来源: AIP
摘要:
A quartz spring balance is described which can be used to measure the magnetic susceptibility of submilligram amounts of sample. The magnetic field is supplied by a moving permanent magnet, and the susceptibility is determined by the deflection of the spring observed in a measuring microscope. The apparatus is calibrated by a comparison standard (platinum) and results are shown for platinum, nickel aluminate, lead, manganese, and sucrose. A precision of better than 2% can be obtained on submilligram amounts of paramagnetic substances having a magnetic susceptibility of from 1 to 50×10−6emu/g. On weakly paramagnetic or diamagnetic substances comparable precision can be obtained on less than 10 mg amounts of sample.