首页   按字顺浏览 期刊浏览 卷期浏览 The sectorial loop‐gap resonator for31P NMR of the adult human liver at 1.5 t with surf...
The sectorial loop‐gap resonator for31P NMR of the adult human liver at 1.5 t with surface tissue suppression

 

作者: A. Jesmanowicz,   W. Froncisz,   Thomas M. Grist,   James S. Hyde,  

 

期刊: Magnetic Resonance in Medicine  (WILEY Available online 1986)
卷期: Volume 3, issue 1  

页码: 76-89

 

ISSN:0740-3194

 

年代: 1986

 

DOI:10.1002/mrm.1910030110

 

出版商: Wiley Subscription Services, Inc., A Wiley Company

 

数据来源: WILEY

 

摘要:

AbstractA new type of loop‐gap resonator, which because of its topology is called a “sectorial resonator,” is described in considerable analytical detail. It was used for31P NMR of the livers of two human adult normal volunteers. An effective strategy for suppression of signals from surface tissue was developed. Depth selection was achieved using a separate excitation coil, permitting a reasonable degree of control of the shape of the sensitive region by adjustment of the dimensions of the excitation coil and of the type of pulse delivered. In addition a new principle for cancellation is described: the topology of the sectorial resonator leads to cancellation signals from tissue masses at the two openings of the structure. Using a phase‐altenated repeated 90° FID pulse sequence, good spectra were achieved in 2.5 min that are free from contamination by the phosphocreatine peak. © 1986 Academic P

 

点击下载:  PDF (751KB)