首页   按字顺浏览 期刊浏览 卷期浏览 Long‐path absorption cell for use with condensable samples
Long‐path absorption cell for use with condensable samples

 

作者: S. L. Bragg,   C. E. Wiswall,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1986)
卷期: Volume 57, issue 1  

页码: 13-16

 

ISSN:0034-6748

 

年代: 1986

 

DOI:10.1063/1.1139136

 

出版商: AIP

 

数据来源: AIP

 

摘要:

A small‐volume long‐path absorption cell has been designed for use with condensable samples. The absorption cell is temperature controlled and operates over the temperature range of −10 to 80 °C. To prevent condensation of the sample on the mirror surfaces, the mirrors are individually temperature controlled and maintained a few degrees above the ambient temperature of the absorption cell. The cell has been tested and used to measure weak absorptions in water vapor in the near infrared.

 