首页   按字顺浏览 期刊浏览 卷期浏览 A Comparative Study of Alternative Bone‐Conduction Calibration Methods
A Comparative Study of Alternative Bone‐Conduction Calibration Methods

 

作者: James Lankford,   Mary Evans,  

 

期刊: Ear and Hearing  (OVID Available online 1981)
卷期: Volume 2, issue 4  

页码: 139-142

 

ISSN:0196-0202

 

年代: 1981

 

出版商: OVID

 

数据来源: OVID

 

摘要:

The accepted instrument for calibrating the bone-conduction section of an audiometer is the artificial mastoid. For a variety of reasons, alternative calibration methods are in general use. Three common methods are: (1) the input voltage method; (2) the real-ear threshold method using normal-hearing listeners; and (3) the real-ear threshold method using subjects with sensorineural hearing loss. The present investigation compared these methods for both accuracy and efficiency.There were no significant differences in accuracy found among the three calibration methods. Substantial differences in efficiency were noted, however. When accuracy and efficiency are considered, the input voltage method appears to be the best alternative method for calibrating the bone-conduction system of an audiometer.

 

点击下载:  PDF (451KB)