An Enzymatic Chemiluminescence Optrode for Choline Detection Under Flow Injection Conditions
作者:
H. Lapp,
U. Spohn,
D. Janasek,
期刊:
Analytical Letters
(Taylor Available online 1996)
卷期:
Volume 29,
issue 1
页码: 1-17
ISSN:0003-2719
年代: 1996
DOI:10.1080/00032719608000387
出版商: Taylor & Francis Group
关键词: Chemiluminescence optrode;choline;bienzyme membrane;cholin;flow cell
数据来源: Taylor
摘要:
A sensitive choline sensor was developed on the basis of a chemiluminometric bienzyme optrode. Choline oxidase (ChOx) and fungal peroxidase (FPOD) were covalently immobilized on a preactivated nylon membrane. The addition of polyethylenimine to the mixed enzyme solution before the immobilization combined with the crosslinking by gaseous glutardialdehyde improves both the long-term stability and the sensitivity. Choline can be detected in the range between 10−6and 10−3M.