首页   按字顺浏览 期刊浏览 卷期浏览 Comparison of a glass thermometer against a nuclear orientation thermometer in high mag...
Comparison of a glass thermometer against a nuclear orientation thermometer in high magnetic fields

 

作者: P. J. Reijntjes,   W. van Rijswijk,   G. A. Vermeulen,   G. Frossati,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1986)
卷期: Volume 57, issue 7  

页码: 1413-1415

 

ISSN:0034-6748

 

年代: 1986

 

DOI:10.1063/1.1138611

 

出版商: AIP

 

数据来源: AIP

 

摘要:

We have compared the low‐frequency capacitance of a glass thermometer made of very pure SiO2doped with ∼1200 ppm OH−with a60CoCosingle‐crystal &ggr;‐ray anisotropy primary thermometer in the range of 15–40 mK in magnetic fields up to 6 T. Within our temperature accuracy of ±5% no shift of the capacitance due to the applied fields was observed.

 

点击下载:  PDF (302KB)



返 回