Extended Angular Range Direct Reading Phase Meter
作者:
S. Bigelow,
J. Wuorinen,
期刊:
Review of Scientific Instruments
(AIP Available online 1957)
卷期:
Volume 28,
issue 9
页码: 713-717
ISSN:0034-6748
年代: 1957
DOI:10.1063/1.1715983
出版商: AIP
数据来源: AIP
摘要:
Most commercial phase meters have an inherent range of 0° to 360°. These instruments usually have a switching arrangement by means of which the range of angles from 360° lagging to 360° leading is covered in a number of sub‐ranges. In the development of a specialized computer, the complex plane scanner, a phase meter was required capable of measuring angles from 540° lagging to 540° leading in one continuous range without ambiguity of output. This paper describes a direct reading pulse position comparison instrument with an inherent range of from 720° lagging to 720° leading. The design is such that this range may be further extended, theoretically without limit. The accuracy of the instrument is ±1° for angles from 540° lagging to 540° leading and for input amplitudes from 0.1 to 100 v. While the phase meter described is designed for operation at a fixed frequency of 400 cps, modifications are suggested for adapting it to broad‐band applications.