首页   按字顺浏览 期刊浏览 卷期浏览 An Automatic Data Acquisition System for Calorimetry: The Specific Heat of &agr;&hyphen...
An Automatic Data Acquisition System for Calorimetry: The Specific Heat of &agr;‐Al2O3between 300 and 470 K

 

作者: Douglas L. Martin,   Robert L. Snowdon,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1970)
卷期: Volume 41, issue 12  

页码: 1869-1876

 

ISSN:0034-6748

 

年代: 1970

 

DOI:10.1063/1.1684432

 

出版商: AIP

 

数据来源: AIP

 

摘要:

The data acquisition system, using a commercially available automatic ac bridge, digital voltmeter, and transistorized digital logic modules, is suitable for both discrete and continuous heating calorimetry. The output, via a Teletype, is a printed record and punched paper tape. The total cost of parts was less than $25 000. The computer program to handle the output is described. As an illustration of the operation of the system the results of measurements in the range 300–470 K on the calorimetric standard sample &agr;‐Al2O3are given. The measurements, having a precision of ±0.1%, were made in an improved form of the calorimeter described by Martin and Snowdon [Can. J. Phys.44, 1449 (1966)]. Agreement with the latest National Bureau of Standards results (Ditmars and Douglas) is around 0.1% when they are corrected to the same temperature scale. The effect of heater lead resistance on the results has been experimentally determined.

 

点击下载:  PDF (619KB)