首页   按字顺浏览 期刊浏览 卷期浏览 An Impedance Method for Blood Pressure Measurement in Awake Rats Without Preheating
An Impedance Method for Blood Pressure Measurement in Awake Rats Without Preheating

 

作者: SUNG-FENG WEN,   JOSEPH TREMBLAY,   MINGHAI QU,   JOHN WEBSTER,  

 

期刊: Hypertension  (OVID Available online 1988)
卷期: Volume 11, issue 4  

页码: 371-375

 

ISSN:0194-911X

 

年代: 1988

 

出版商: OVID

 

关键词: tail-cuff method;temperature effect;hypertension;diabetes mellitus

 

数据来源: OVID

 

摘要:

The tail-cuff methods for measuring systolic blood pressure in the rat usually require preheating of the animal to obtain recordable pulse signals. To find a more sensitive method, we applied the principle of differentiated impedance (dZ/dt) to the tail-cuff measurement of systolic blood pressure. We obtained clear pulse signals from the tail in awake rats without preheating the animals, and the systolic blood pressure obtained by this method had an excellent correlation with the directly measured femoral artery pressure (correlation coefficient=0.98). Heating the animals at 40°C for 5 minutes increased systolic blood pressure by a mean of 6 mm Hg as compared with that determined at the ambient temperature of 21 to 24°C. Mean systolic blood pressure in young female diabetic rats was 122 ± 3 mm Hg, which was significantly higher than the 111 ± 2 mm Hg of normal rats. It is concluded that the technique of electrical impedance as applied to the tail-cuff method is simple and highly sensitive and is suitable for measurement of tail systolic blood pressure in awake rats without preheating.

 