首页   按字顺浏览 期刊浏览 卷期浏览 Apparatus Used in Pulse‐Type Ultrasonic Flaw Detection
Apparatus Used in Pulse‐Type Ultrasonic Flaw Detection

 

作者: Eiji Yamamoto,  

 

期刊: The Journal of the Acoustical Society of America  (AIP Available online 1971)
卷期: Volume 49, issue 6A  

页码: 1709-1716

 

ISSN:0001-4966

 

年代: 1971

 

DOI:10.1121/1.1912566

 

出版商: Acoustical Society of America

 

数据来源: AIP

 

摘要:

In pulse‐type ultrasonic testing, the size of the defect or flaw in a specimen is estimated by the size of the echo reflected by the defect of flaw. By this method, the acceptability of the specimen is determined. In today's ultrasonic flaw detecting, the reflection coefficient of flaws, the characteristics of the acoustic field, the echoes from flaws in specimens, etc., are handled as those of continuous waves. However, in order to improve the resolving power, which is one of the flaw‐detecting capabilities of the ultrasonic flaw‐detecting method, it is necessary to minimize the number of waves contained in the pulses, and because of this the frequency distribution becomes wide. Under a condition such as this, it is not possible to determine the quantitativeness of echo from a flaw. With respect to the above problem, the author conducted a study especially on the probe, transmitter, receiver amplifier, etc., of the apparatus used in pulse‐type ultrasonic testing, mainly in order to obtain the quantitativeness of the echoes reflected by flaws.

 

点击下载:  PDF (734KB)



返 回